| Detection
method |
Absorbance |
| Wavelength
range (nm) |
200
- 999 |
| Wavelength
accuracy (nm) |
±2.0 |
| Wavelength
repeatability (nm) |
±0.2 |
| Wavelength
selection |
Monochromator
(tunable in 1nm increments) |
Photometric
accuracy - typical (%) |
<1.0
± 0.01Abs |
| Photometric
linearity - typical (%) |
±1.0 |
| Photometric
repeatability - typical (%) |
0.5
± 0.005Abs |
| Photometric
range - typical (Abs) |
0
- 4.0 |
| Stray
light - typical at 230nm (%) |
0.03 |
| Bandpass
(nm) |
5.0 |
| Read
method |
Endpoint,
Kinetic, Spectral scanning and Well area scanning |
Reading
Speed 96 wells - Normal mode (sec) |
20 |
Reading
Speed 384 wells - Normal mode (sec) |
66 |
Reading
Speed 96 wells - Rapid mode (sec) |
11 |
Reading
Speed 384 wells - Rapid mode (sec) |
23 |
Reading
Speed 96 wells - Sweep mode (sec) |
5 |
Reading
Speed 384 wells - Sweep mode (sec) |
11 |
| Temperature
control |
4°
above ambient to 50°C |
| Shaking |
User
programmable frequency and duration |
| Control
and data analysis software |
Gen5
included |
| Automation |
Robot
accessable plate carrier - Bio-Stack compatible |